2. High-Voltage Adjustment (SN 3001A and Below)
10. Wait at least one minute for capacitors to discharge to a safe
level.
11. Remove the protective cover from the
High-Voltage
Regulator. A label should be visible on the
High-Voltage
Transformer. Record the voltage listed on the label for use in step
15.
If the label is missing, use the nominal value of -3790 Vdc.
12. Connect the high-voltage probe to
See Figure 3-7 for
the location of the test point.
With power supplied to the instrument,
is at a voltage
level of approximately -4000 V dc. Be extremely careful.
H i g h V o l t a g e
R e g u
L o b e I
H V T P
Figure 3-7. Location of Label and
Point
13. Reconnect ac line cords to both instrument sections. Set the
spectrum analyzer LINE switch to ON.
14. Wait approximately 30 seconds for the dc regulator circuits to
stabilize.
15. Adjust
HV ADJ for a DVM indication equal to the
calibration factor (calculated in step 8) times the voltage labeled
on the top of
High-Voltage Regulator (noted in step 11). See
Figure 3-6 for the location of the adjustment.
V dc
E
XAMPLE
:
If the calibration factor calculated in step 8 is 0.00099, and
is labeled for -3875 V, then adjust
HV ADJ for
a DVM indication of:
0.00099 x (-3875 V) = -3.836 V dc
16. With the front-panel INTENSITY control fully counterclockwise,
wait approximately 30 minutes to allow the high-voltage supply to
3-34 Adjustments