3. Preliminary Display Adjustments (SN 3004A and Above)
29. Set the oscilloscope controls as follows:
Press
Channel 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
amplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Press (SHOW].
30. Adjust the spectrum analyzer front-panel INTENSITY control for
50 V peak-to-peak (6.25 divisions) as indicated on the oscilloscope.
See Figure 3-29.
r u n n i n g
- 1 2 5 . 0 0 0 “ s
1 2 5 . 0 0 0
3 7 5 . 0 0 0
5 0 . 0
Figure 3-29.
Signal
1
8 . 0 0
4 5 . 0 0 v
10.00 1
50.00 v
31. Adjust
HF GAIN and
for minimum
overshoot on rise and minimum rise and fall times of the pulse
waveform.
32. Use the oscilloscope
markers to measure the risetime,
falltime, and percent overshoot of the pulse waveform. Rise
and falltimes should be less than 50 ns and 90 ns respectively.
Overshoot on the rise should be less than 5% (approximately 0.4
divisions).
33. Set the spectrum analyzer LINE switch to STANDBY.
34. Disconnect the oscilloscope channel 1 probe from the spectrum
analyzer. Remove the Display Adjustment PC board from the
slot, and reinstall the
Intensity Control Assembly.
Replace the A3 Section cover and cables.
35. Reconnect the black connector with three wires (8, 98, and 96) to
and set
INT GAIN approximately two-thirds
clockwise.
36. Perform Adjustment Procedure 4 Final Display Adjustment (SN
and Above).
Adjustments