![HP HP 8566B Adjustment Manual Download Page 375](http://html.mh-extra.com/html/hp/hp-8566b/hp-8566b_adjustment-manual_161563375.webp)
3 MHz Bandwidth Filter Adjustments
21.4 MHz Bandwidth Filter Adjustments
Down/Up Converter Adjustments
f2.0
10 Hz bandwidth
30 Hz bandwidth
f0.5
100 Hz to 1 MHz bandwidth
fl.O
3 MHz bandwidth
30
and 100
bandwidth switching uncertainty figures only
applicable
Relative Humidity.
The CAL OUTPUT signal is applied to the input of the spectrum
analyzer. The deviation in peak amplitude of the signal trace is then
measured as each resolution bandwidth filter is switched in.
None required
1. Press
2. Connect CAL OUTPUT to RF INPUT.
3. Key in the following control settings:
FREQUENCY)
[
F R E Q U E N C Y
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
REFERENCE LEVEL) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. - 8
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
4. Press LOG
[
ENTER
and key in 1
Press MARKER
(
PEAK
SEARCH)
MHz
MHz
MHz
5. Key in settings according to
4-6. Press MARKER
[
PEAK SEARCH
)
at each setting, then read the amplitude deviation
from the MARKER La] readout at the upper right of the display
(see Figure
allowable deviation for each resolution
bandwidth setting is shown in the table.
Option 462