17.
Loop Phase Lock Adjustments
19. Set the RF Section LINE switch to STANDBY. Disconnect the DVM
from
and the second spectrum analyzer from
OUT 160-166 MHz.
20. Reinstall
PLL3 Up Converter Assembly in the RF Section.
Reconnect the cable 7 (violet) to
SCAN
MHz OUT,
the cable 6 (blue) to
OUT 160-166 MHz, and the cable 8
(gray) to
IN 100 MHz.
Adjustments 3-159