16. Harmonic and Intermodulation Distortion
49. Press MARKER
Activate
[
CENTER FREQUENCY
)
and press
once to tune to the third order product at 3998.5 MHz. Wait for
completion of the sweep.
50. Press MARKER
[
PEAK SEARCH
)
Record the MKR A amplitude:
51. Press
three times to tune to the third order product at 4001.5
MHz. Wait for completion of the sweep, then press MARKER
(
PEAK
SEARCH].
Record the MKR A amplitude:
52. Choose the smallest MKR A amplitude in steps 50 and 51.
Record its absolute value: S =
53. Compute the
TO1 = P +
-25
+
54. The result should be + 7
TO1 for signals of 3999.5 MHz and 4000.5 MHz
55. Repeat steps 41 through 53 for the input signal frequencies and
the third order products shown in
2-19. The TO1 for each
setting should be + 5
2-19. TO1 Measurement Settings
Input Signal Frequencies Third Order Products Third Order Suppression
8999.500
9000.500
8998.500
9001.500
13999.500 14000.499
13998.501 14001.498
2-72 Performance Tests