Keyboard
X, Y, Z Axis Amplifier
Adjustment Procedure 2, “High-Voltage Adjustment,” should be
performed before performing the following adjustment procedure.
Perform this adjustment only if components have been replaced on
the
X, Z Axis Amplifier Assembly. Components R117, R217,
R308,
and
are factory adjusted and
normally do
not
require readjustment. Components affecting these
adjustments are located in function blocks F, H, M, N, 0, P, R, and S
of the
X, Y, Z Axis Amplifier Assembly schematic diagram.
The X, Z Axis Amplifier Assembly is adjusted to compensate the
CRT drive circuits for proper horizontal and vertical characteristics.
These preliminary adjustments are necessary only after replacement
or repair of the
X, Y, Z Axis Amplifier Assembly). For routine
maintenance, CRT replacement, or minor repairs, only Adjustment
Procedure 4, “Final Display Adjustments,” needs to be performed.
Be sure not to allow a fixed spot of high intensity to remain on the
spectrum analyzer CRT. A high intensity spot may permanently
damage the CRT’s phosphor coating. Monitor the CRT closely during
the following adjustment procedures. If a spot occurs, move it
off-screen by adjusting either the front-panel INTENSITY control, or
the horizontal or vertical deflection position controls.
Digitizing Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Pulse/Function Generator . . . . . . . . . . . . . . . . . . . . . . . . . HP 8116A
10: 1 Divider Probe, 10
(2
required) . . . . .
HP
Display Adjustment PC Board
accessory) . . . . .
Termination, BNC
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 11593A
Adapters:
Adapter, BNC(f) to
. . . . . . . . . . . . . . . . . . . . . . . .
Adapter, BNC tee . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-56 Adjustments