2 1. Frequency Response Adjustments
73. On
the
spectrum analyzer, key in
TIME
)
LOG SCALE
[
ENTER
1
TRACE B
4,
TRACE B
74. Readjust 17-turn (YTX Diode Bias) potentiometer
VB
to maximize the overall level of the displayed signal responses
from 2.0
to 5.8
on the spectrum analyzer display.
75. Readjust 25-turn (YTX Drive) potentiometers
5.8
and
2
to maximize the overall level of
the displayed signal responses from 2.0
to 5.8
on
the
spectrum analyzer display.
76. Gradually readjust (Band B Lower Segment)
and (Band B Upper Segment)
B2 as necessary for
maximum flatness of the displayed signal responses.
77. On the spectrum analyzer, key in SWEEP
&
WEEP TIME
)
TRACE A
TRACE B
6, TRACE B
(
CLEAR
-
WRITE
],
78. Press SWEEP
on
the spectrum analyzer and wait for the
sweep to complete (150 seconds) and the SWEEP LED to turn
off. As the spectrum analyzer tunes from 2.0
to 5.8
the spectrum analyzer frequency response should be displayed as
TRACE B, as shown in Figure 3-96.
S T
Figure 3-96.
Typical Frequency Response (2.0
5.8
79. On the spectrum analyzer, key in (RECALL) 4, (HOLD) and repeat
steps 74 through 76 as necessary.
80. On the spectrum analyzer, key in [RECALL) 5 and wait for the
sweep to complete (150 seconds) and the SWEEP LED to turn
Off.
3-188 Adjustments