8. 21.4 MHz Bandwidth Filter Adjustments
43. Connect
CAL OUTPUT to RF INPUT through 1
and 10
step
attenuators. Set step attenuators to 25
4 4 .
Key in
(
CENTER FREQUENCY
_)
100 MHz,
(
FREQUENCY
3
0
1
z, (
REFERENCE LEVEL
)
-30
45.
Key in LOG
C
ENTER
1
MARKER [al
46.
Key in
(
REFERENCE LEVEL
]
-20
Set step attenuators to 15
47. Adjust
to align markers on display. MARKER
A level should indicate 0.00
See Figure 3-43 for location of
adjustment.
48.
Key in
[
REFERENCE LEVEL
]
-10
Set step attenuators to 5
49. Adjust
to align markers on display. MARKER
A level should indicate 0.00
See Figure 3-43 for location of
adjustment.
50. Refer to Performance Test 5, “Resolution Bandwidth Switching
Uncertainty Test”, and check the amplitudes of resolution
bandwidths from 1
through 3 MHz. If the amplitude of the
300
resolution bandwidth is more than 0.3
low relative to
the 100
and 1 MHz resolution bandwidths, perform steps 51
through 71, LC Dip Adjustments.
If the amplitudes of the 3
10
and 30
resolution
bandwidths are not within
of the amplitude of the 1
MHz resolution bandwidth, perform steps 1 through 12 (Bandpass
Filter Adjustments) of Adjustment Procedure 5, “Log Amplifier
Adjustments” and then check the amplitudes of the resolution
bandwidths from 3
through 3 MHz again. If the amplitudes
of the 3
10
and 30
resolution bandwidths are still
not within f0.4
of the 1 MHz resolution bandwidth, change
the value of factory-select component
or
An increase of one standard value of
decreases the
amplitudes of the 100
through 3 MHz resolution bandwidths
by approximately 0.15
An increase of one standard value of
decreases the amplitudes of the 3
through 30
resolution bandwidths by approximately 0.05
If the amplitudes of the 3
through 3 MHz resolution
bandwidths are not within
of
the
amplitude of the 1
resolution bandwidth, perform Adjustment Procedure 7, “3 MHz
Bandwidth Filter Adjustments, and Adjustment Procedure 11,
“Down/Up Converter Adjustments.
51. Set spectrum analyzer LINE switch to STANDBY.
52. Disconnect cable 97 (white/violet) from
and connect cable
to
53. Remove
Bandwidth Filter and install on 2 extender boards.
54. Set spectrum analyzer LINE switch to ON. Press
5 5 .
Key in
(
CENTER FREQUENCY
]
100 MHz,
100
1 MHz,
0
LOG
[
ENTER
56. Set step attenuators to 0
Short
to ground.
Adjustments 3-67