12. Noise Sidebands
SPECTRUM ANALYZER
ADAPTER
ADAPTER
Figure 2-22. Noise Sidebands Test Setup
Spectrum Analyzer
LO OUTPUT) . . . . . . . . . . . . . . . . . . . HP
Adapter, Type N (m) to SMA (f) (2 required) . . . . . . . . . . . . . .
Adapter, Type N (f) to BNC (m) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1250-0077
Cable Assembly, SMA Male Connectors . . . . . . . . . . . . . . . . . . .
1. Allow both analyzers to warm up for at least one-half hour with
LINE switch in either the STANDBY or ON position.
2. On one of the analyzers, connect the CAL OUTPUT to the RF
INPUT. Press
then
and adjust FREQ ZERO
control for maximum signal amplitude. Repeat this frequency
calibration on the other analyzer. When completed, press
again on each of the analyzers.
3. Connect
LO OUTPUT of source analyzer to RF INPUT of
analyzer under test as shown in Figure 2-22.
Do not connect the frequency reference (on the rear panel) of the
analyzers to a common frequency reference.
4. Key in the following on the source analyzer:
(
CENTER
FREQUENCY) . . . . . . . . . . . . . . . . . . . .
(
FREQUENCY SPAN
) . . . . . . . . . . . . .
SHIFT MKR REF LVL
(Display diagnostics
[SHIFT]
(YTO Pretest Mode)
SWEEP
5.7
. . . . . . . . . . . . . . . .
for convenience)
The first line of the diagnostic display and the CENTER readout
should both now indicate 5.700 000 000
This is the
LO
OUTPUT frequency.
Performance Tests