14. Average Noise Level
8.
9.
10.
11.
Figure 2-26. Average Noise Level Measurement
Read the noise level from the MARKER amplitude readout.
value should be less than -95
Key in the following settings:
The
51
.
20
ms
Wait several seconds for the trace to stabilize (VID AVG
Read the amplitude from the MARKER readout. The value should
be less than -112
Set
(
CENTER FREQUENCY
)
according to
2-18. At each setting
allow several seconds for the trace to stabilize before reading the
amplitude from the MARKER readout. The maximum allowable
level for each frequency is given in the table.
Performance Tests 2-59