10. Frequency Response Test
2-15. Frequency Response (Flatness)
1
2
3
4
y
Spectrum Analyzer
C d
Trace Limits
Rand
and
y
Flatness
Synthesized Sweeper
Minimum
Maximum
START
STOP
Power
Amplitude
y Amplitude
y
FREQ
FREQ
Sensor
MHz 2.5
60 MHz
2.5
100 MHz
-11.20
-8.80
1.20
2 5.8
2
3.9
3
3.9
5.8
5
12.30
-7.70
3.40
5 . 8 1 2 . 5
5 . 8
9.15
7
9.15
12.5
11
12.30
-7.70
3.40
12.5 18.6
12.5
15.55
14
15.55
18.6
17
-12.80
-7.20
4.40
18.6 20
18.6
20
19
12.80
-7.20
4.40
20 22
20
22
21
-13.60
-6.40
6.00
umulative Flatness
100 Hz to 20
Specification: 4.40
100 Hz to 22
Specification: 6.00
Performance Tests 2-45