3. Preliminary Display Adjustments (SN 3001A and Below)
The pulse/function generator output must be terminated with 50
ohms.
Use a BNC tee, a
termination, and a BNC female to SMB
adapter. Install the
termination as close to
the
Display
Adjustment PC Board as possible.
26. Set the pulse/function generator controls as follows:
MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . NORM
Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Frequency (FRQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Width (WID) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250 ns
Amplitude (AMP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Offset (OFS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
27. Set the spectrum analyzer front-panel INTENSITY control fully
clockwise. Adjust the oscilloscope trigger level for a stable display.
Note the display on the oscilloscope. The pulse should be
peak-to-peak.
28. Set the oscilloscope controls as follows:
Press
Channel 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
amplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...8.0
Press
29. Adjust
X POS and
Y POS to either extreme to
position the CRT beam off-screen (to prevent possible damage to
the CRT phosphor). If it is not sealed, adjust
INT GAIN
fully clockwise.
30. Adjust the spectrum analyzer front-panel INTENSITY control for
50 V peak-to-peak (6.25 divisions) as indicated
on
the oscilloscope.
See Figure 3-22.
- 1 2 5 . 0 0 0
1 2 5 . 0 0 0 “ s
5 0 . 0
1
8 . 0 0
4 5 . 0 0 v
10.00 dc
50.00 v
Figure 3-22.
Signal
31. Adjust
HF GAIN and
for minimum overshoot
on rise and minimum rise and fall times of the pulse waveform.
3-54 Adjustments