16. YTO Loop Adjustments
Frequency determined in step 25, and set the power meter CAL
FACTOR control accordingly. Connect the power sensor to the
SMA female end of W 11 (W 11 should still be connected to the
COUPLED OUT connector of
CIA Assembly).
28. Adjust
GATE BIAS ADJ for a power meter indication of
-5.0
fO.l
29. Disconnect the power sensor from semi-rigid coax cable
Disconnect semi-rigid coax cable W 11 from the COUPLED OUT
connector of
CIA Assembly, and reconnect
to the
CIA Assembly.
30. Remove the cover from
YTO Loop Sampler Assembly.
Locate DRIVER MATCHING adjustments Al
and Al
SAMPLER DC test point
and IF GAIN adjustment
These parts are indicated on the cover of the
YTO Loop Sampler Assembly and in Figure 3-73.
I M P E D A N C E
M A T C H
C l
c22
I F G A I N
Figure 3-73. Al
Adjustment Locations
31. On the oscilloscope, key in
to perform a soft
reset.
3-140 Adjustments