Technical Service Manual
A - 1
430-00587-008 (Rev. 1/05)
CONSIDERATIONS FOR USE IN
ELECTROMAGNETIC ENVIRONMENTS
The LifeCare XLM with DataPort (outside of the United States) is intended for use
in the electromagnetic environment specified in
Electromagnetic Immunity for Life-Supporting Equipment
. The user of the device should assure that it is used only in the appropriate
environment.
Note:
Do not connect DataPort when infusing.
Table A-1, Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
electromagnetic emissions compliance and guidance for XLM Series infusion systems
Table A-1.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic Enforcement - Guidance
RF Emissions
CISPR 11
Group 1
The LifeCare XLM with DataPort uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The LifeCare XLM with DataPort is suitable for use
in all establishments, including domestic
establishments and those directly connected to
the public low voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class B
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
EN-2