Technical Service Manual
5 - 1
430-00587-008 (Rev. 2/05)
A complete maintenance program promotes infusion longevity and trouble-free instrument
operation. Such a program should include routine maintenance, periodic maintenance
inspection, and following any repair procedure, performance verification testing.
Routine maintenance consists of basic inspection and cleaning procedures. As a minimum
requirement, inspect and clean the infusion pump after each use. In addition, establish
a regular cleaning schedule for the infusion pump.
INSPECTING THE INFUSION SYSTEM
Inspect the infusion system periodically for signs of defects such as worn accessories,
broken instrument connections, or damaged cables. In addition, inspect the infusion
system after repair or during cleaning. Replace any damaged or defective external parts.
See
The following procedures are designed to maintain the infusion pump, sustain system
longevity, and promote trouble-free instrument operation.
Follow hospital protocol for establishing the infusion pump cleaning schedule.
CAUTION:
Do not immerse the infusion system in liquids. Immersion could damage
the instrument. Do not allow liquids to enter the infusion system electronics
compartment. Do not spray cleaning solutions toward any openings in the infusion
system.
WARNING:
DISCONNECT THE INFUSION SYSTEM FROM AC (MAINS)
POWER PRIOR TO CLEANING THE INSTRUMENT. FAILURE
TO COMPLY WITH THIS WARNING COULD RESULT IN
ELECTRICAL SHOCK.