APPENDIX
430-00587-008 (Rev. 1/05)
A - 4
Plum XL Series
Note:
At 80 MHz and 800 MHz, the higher frequency range applies.
Note:
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
a
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are
6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and
40.660 MHz to 40.700 MHz.
b
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in
the frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood that mobile/
portable communications equipment could cause interference if it is inadvertently brought
into patient areas. For this reason, an additional factor of 10/3 is used in calculating the
recommended separation distance for transmitters in these frequency ranges.
c
Field strengths from fixed transmitters, such as base stations for radio (cellular and/or
cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the LifeCare
XLM with DataPort is used exceeds the applicable RF compliance level above, the LifeCare
XLM with DataPort should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or
relocating the LifeCare XLM with DataPort.
d
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V
1
] V/m.
Radiated RF
IEC
61000-4-3
10V/m
80 MHZ to
2.5 GHz
[E
1
] V/m
Recommended separation distance:
Where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation distance in
meters (m).
b
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
C
should be
less than the compliance level in each
frequency range.
d
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Table A-3.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity for
Life-Supporting Equipment and Systems
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Immunity-Guidance
d
12
E
1
------
P
80 MHz to 800 MHz
=
d
23
E
1
------
P
800 MHz to 2,5 GHz
=