![Hospira LifeCare XL Series Technical & Service Manual Download Page 71](http://html1.mh-extra.com/html/hospira/lifecare-xl-series/lifecare-xl-series_technical-and-service-manual_2165611071.webp)
SECTION 6 TROUBLESHOOTING
430-00587-008 (Rev. 2/05)
6 - 8
Plum XL Series
60-1
Plunger motor position
flag stuck high during
re-synchronization
Position sensor failure
Replace mechanism
assembly
Plunger motor not moving
Check connection
at J7 on MCU PWA
If condition recurs,
60-2
Plunger motor position
signal is continuous
low during
re-synchronization
Position sensor failure
Replace mechanism
assembly
Enable circuit failed
Check connection
at J7 on MCU PWA
If condition recurs,
Plunger motor not moving
61-1
Input/output motor
position flag is
continuous high during
re-synchronization
Position sensor failure
Replace mechanism
assembly
Input/output motor
not moving
Check connection
at J8 on MCU PWA
If condition recurs,
61-2
Input/output motor
position signal is
continuous low during
re-synchronization
Position sensor failure
Replace mechanism
assembly
Enable circuit failure
Check connection
at J8 on MCU PWA
If condition recurs,
Input/output motor
not moving
62-1
Primary/secondary
motor position flag is
continuous high during
re-synchronization
Position sensor failure
Replace mechanism
assembly
Primary/secondary
motor not moving
Check connection
at J9 on MCU PWA
If condition recurs,
62-2
Primary/secondary
motor position signal is
continuous low during
re-synchronization
Position sensor failure
Replace mechanism
assembly
Enable circuit failure
Check connection
at J9 on MCU PWA
If condition recurs,
Primary/secondary
motor not moving
63-1
Plunger motor phase loss
Plunger motor does not
have enough torque
Replace mechanism
assembly
Mechanical assembly
failure
Table 6-2.
Error Codes Requiring Technical Service
Error
Code
Malfunction
Description
Possible Cause