5.5
DATAPORT CONNECTION AND GROUND CONTINUITY TEST (OPTIONAL)
Technical Service Manual
5 - 17
430-00587-008 (Rev. 2/05)
8. Set the terminal emulator in the host computer with the following parameters:
- 1200 baud, 8 data bits, parity none, stop bits 1
Echo typed character locally
9. Turn the control knob to the SET RATE position.
10. Type in the following commands from the terminal emulator:
T@0;I45A4<CR> (where “<CR>” = Carriage Return)
Note:
Commands are case sensitive.
11. Verify the response message is in the following format:
FXX;YYYY;RZZZZ (where “XX = Hard ID”, “YYYY = Soft ID”, “ZZZZ = CRC”)
12. Turn the control knob to the OFF position.
13. Disconnect the infusion system from the RS-232 cable DataPort Connector.
Note:
Do not connect DataPort when infusing.
Figure 5-8.
DataPort Connector
PIN 10 (CTS)
PIN 9 (COMGND)
1
AUDIBLE
LEVEL SWITCH
REAR
ENCLOSURE
8
04K01029
PIN 4 (NURSE1)
PIN 5 (NURSE2)