1.8
INSTRUMENT INSTALLATION PROCEDURE
Technical Service Manual
1 - 9
430-00587-008 (Rev. 2/05)
Inspect the infusion system for shipping damage. Should any damage be found, contact
the delivering carrier immediately.
CAUTION:
Do not use the infusion system if it appears to be damaged. Should
damage be found, contact Hospira (
see Section 6.1, Technical Assistance
). Do not use
the infusion system if it appears to be damaged.
Inspect the infusion system periodically for signs of defects such as worn accessories,
broken connections, or damaged cable assemblies. Also inspect the infusion system after
repair or during cleaning. Replace any damaged or defective external parts.
CAUTION:
Do not place the infusion system in service if the self test fails.
To perform the self test, refer to
Figure 1-3, Plum XL LCD Test Screens
, and proceed as follows:
1. Connect the infusion system AC (mains) power cord to a grounded AC (mains) outlet
and confirm the AC (mains) power icon (next to the OFF/CHARGE setting)
illuminates.
2. Open the door assembly (cassette door) by lifting up on the cassette door handle.
3. Hold a primed cassette by its handle and insert the cassette into the cassette door
guides. Do not force the cassette into position.
4. Close the cassette door handle to lock the cassette in place.
5. Turn the control knob to SET RATE to initiate the self test.
6. Verify the following screens display in succession:
- LCD test screen
- Four backward Cs (approximately two seconds)
- Set rate screen
7. Disconnect the infusion system from AC (mains) power and confirm BATTERY
displays on the LCD screen.
8. Turn the control knob to OFF/CHARGE and remove the administration set.
9. To allow the battery to charge fully, connect the infusion system to AC (mains) power
for a minimum of eight hours with the control knob in the OFF/CHARGE position.
Confirm the AC (mains) power icon illuminates.