SECTION 6 TROUBLESHOOTING
430-00587-008 (Rev. 2/05)
6 - 4
Plum XL Series
14-1
LOCKED
Lock
violation
alarm
Control knob
position changed
while in LOCKED
mode
Press LOCK
button and reset
settings
15-1
None
The device does
not communicate
to the host computer
UART test failure
The UART
loop-back test
during power-up
failed
Replace
MCU PWA
15-2
None
The device does
not communicate
to the host computer
Excessively
frequent UART
interrupts
The MCU does
not have enough
time to process
UART hardware
due to UART
hardware failure
Replace
MCU PWA
15-3
None
The device does
not communicate
to the host computer
The UART
receiver buffer
has overflowed
The MCU does
not have enough
time to process
the received data
due to UART
hardware failure
Replace
MCU PWA
16-1
TURN TO RUN
Control knob in
between valid
states for five
minutes
Control knob not
in OFF/CHARGE
or RUN position
Turn control
knob to RUN,
OFF/CHARGE,
or HOLD/RESET
position
Defective control
knob assembly
Replace control
knob assembly
17-1
LOW BATTERY
Low
battery
alarm
Low battery
Connect to AC
power or turn
control knob to
HOLD/RESET,
then to RUN
17-2
Low battery
re-alarms after
15 minutes
Low battery
Connect to AC
power or turn
control knob to
HOLD/RESET,
then to RUN
18-1
LOW BATTERY
Discharged
battery alarm
Fully
discharged
battery
Connect to AC
power, turn
control knob to
OFF/CHARGE,
or replace
the battery
Table 6-1.
Operational Alarm Messages and Corrective Actions
Error
Code
Alarm Message
Description
Possible Cause
Corrective
Action