7.2
REPLACEMENT PROCEDURES
Technical Service Manual
7 - 35
430-00587-008 (Rev. 2/05)
Figure 7-17.
Top View of the Main Chassis Assembly
MECHANISM ASSEMBLY REPLACEMENT
The recommended tool for this procedure are diagonal cutters and a medium size flat blade
screwdriver.
The replacement parts for this procedure are:
Assembly, Mechanism
Screw, 6-32 x 1/2, Hex Head, Slotted
Tie, Cable
Mount, Cable Tie, Adhesive-Back
CAUTION:
Use proper ESD grounding techniques when handling components.
Wear an antistatic wrist strap and use an ESD-protected workstation. Store the PWA
in an antistatic bag before placing it on any surface.
To replace the mechanism assembly, refer to
, then proceed as follows:
1. Separate the front enclosure assembly, rear enclosure assembly, and main chassis
assembly as described in
MCU PWA
FERRITE
P5
J5
SENSOR/MCU
CABLE
BATTERY
COMPARTMENT
04K01028
Summary of Contents for LifeCare XL Series
Page 5: ...CHANGE HISTORY 430 00587 008 Rev 2 05 ii Plum XL Series This page intentionally left blank...
Page 271: ...APPENDIX 430 00587 008 Rev 1 05 A 6 Plum XL Series This page intentionally left blank...
Page 277: ...INDEX 430 00587 008 Rev 2 05 I 6 Plum XL Series This page intentionally left blank...