6.2
ALARM MESSAGES AND ERROR CODES
Technical Service Manual
6 - 7
430-00587-008 (Rev. 2/05)
45-1
[PRI/SEC] key stuck
in ON position
Switch S1 on display PWA
shorted or stuck
Replace display PWA
Front panel key stuck
Replace key
Replace front
enclosure assembly
45-2
[UP ARROW] key stuck
in ON position
Switch S3 on display PWA
shorted or stuck
Replace display PWA
Front panel key stuck
Replace key
Replace front
enclosure assembly
45-3
[DOWN ARROW] key
stuck in ON position
Switch S2 on display PWA
shorted or stuck
Replace display PWA
Front panel key stuck
Replace key
Replace front
enclosure assembly
45-4
[TITRATE] key stuck
in ON position
Switch S4 on display PWA
shorted or stuck
Replace display PWA
Front panel key stuck
Replace key
Replace front
enclosure assembly
45-5
[BACKPRIME] key stuck
in ON position
Switch S5 on display PWA
shorted or stuck
Replace display PWA
Front panel key stuck
Replace key
Replace front
enclosure assembly
45-6
[SILENCE] key stuck
in ON position
Switch S6 on display PWA
shorted or stuck
Replace display PWA
Front panel key stuck
Replace key
Replace front
enclosure assembly
59-1
Valve motor moving
at the wrong time
Position sensor failure
Replace mechanism
assembly
Motor drive circuit failure
Replace MCU PWA
Table 6-2.
Error Codes Requiring Technical Service
Error
Code
Malfunction
Description
Possible Cause
Corrective Action