7.2
REPLACEMENT PROCEDURES
Technical Service Manual
7 - 15
430-00587-008 (Rev. 2/05)
To replace the rear enclosure assembly components, proceed as detailed in the following
sections.
Recommended tools for this procedure are a medium size flat blade screwdriver and a fuse
puller.
The replacement part for this procedure is:
Fuse, 0.5A, 250V, Slo-Blo, 5 x 20 mm
Cover, Fuse 5 x 20 mm
Figure 7-5, Fuse Replacement (115 VAC)
, then proceed as follows:
1. Separate the front enclosure assembly and rear enclosure assembly as described
2. Remove the two plastic fuse covers. Using the fuse puller, remove the two fuses on
the power supply PWA and install replacements.
CAUTION:
Confirm replacement fuse rating is identical to fuse rating
indicated on the power supply PWA or equipment damage may occur.
3. Reassemble the device in the exact reverse order of disassembly.
Note:
(XLM with DataPort) When joining the front and rear enclosures, do not
pinch the DataPort/MCU cable between the two enclosures.
4. Connect the infusion system to AC (mains) power.