6.2
ALARM MESSAGES AND ERROR CODES
Technical Service Manual
6 - 3
430-00587-008 (Rev. 2/05)
08-1
AIR IN LINE
BACKPRIMING
Empty
container
alarm
500 µL of air
detected entering
the cassette in
the last two intake
strokes
Change
container
and backprime
to expel air
10-1
CHECK SETTINGS
mL/H=mL?
Check
settings
alarm
Rate and VTBI
settings not
correct
Turn control
knob to
SET RATE
or SET VTBI
to check settings
or enter values
11-1
TURN TO RUN
Turn to run alarm
Rotary control
knob not in
OFF/CHARGE
or RUN position,
or no key is
pressed for
five minutes
Turn control
knob to RUN,
OFF/CHARGE,
or HOLD/RESET
12-1
VTBI COMPLETE
mL=mL
Primary VTBI
complete alarm
The VTBI for
the primary
channel has
been delivered
Discontinue
infusion,
or change
container and
program new
VTBI setting
12-2
Secondary VTBI
complete alarm
The VTBI for
the secondary
channel has
been delivered
13-1
CASSETTE
DOOR/CASSETTE
(XL only)
or
Input/output
valve leak
test failure
Defective
administration set
Cassette
improperly
loaded or
improperly
primed
Turn control
knob to
OFF/CHARGE,
open and close
cassette door,
then restart
If condition
recurs, replace
administration
set
Turn control
knob to OFF/
CHARGE or
HOLD/RESET,
reprime cassette
and restart
13-2
Primary/
secondary
valve leak
test failure
13-3
Valve leak
test failure
due to excessive
signal noise
Fluid spillage
around valve pins
Clean valve pins
Table 6-1.
Operational Alarm Messages and Corrective Actions
Error
Code
Alarm Message
Description
Possible Cause
Corrective
Action