SECTION 7 REPLACEABLE PARTS AND REPAIRS
430-00587-008 (Rev. 2/05)
7 - 4
Plum XL Series
Recommended tools and materials for this procedure are a 3/8 inch wood chisel or an
X-acto knife and mild solvent.
The replacement part for this procedure is:
Foot Pad, Rubber
To replace the rubber foot pads, refer to
Figure 7-1, Bottom View of the Infusion System
then proceed as follows:
1. Turn the control knob to OFF/CHARGE.
2. Disconnect the infusion system from AC (mains) power.
3. Set the infusion system on its back to access the foot pads.
4. Using a 3/8 inch wood chisel or an X-acto knife, remove the rubber foot pad and
scrape the enclosure recess to remove adhesive residue.
Note:
Each adhesive-backed rubber foot pad is bonded in its recess. Do not
damage the recess.
5. Using a mild solvent, clean the enclosure recess.
6. Remove the protective backing from the self-adhesive surface on the replacement
rubber foot pad and bond the foot pad in place.
7. After approximately five minutes, verify the foot pad is secure.
8. Connect the infusion system to AC (mains) power.
Replacement of a rubber foot pad is a routine maintenance procedure and no verification
procedure is normally required. However, if the infusion system may have been damaged
during this procedure, perform the PVT in