7.2
REPLACEMENT PROCEDURES
Technical Service Manual
7 - 23
430-00587-008 (Rev. 2/05)
Figure 7-10.
AC Power Cord (220 VAC - Plum XLM)
LIFECARE AC RECEPTACLE, BRACKET, GASKET, SEAL, AND
EQUIPOTENTIAL POST REPLACEMENT
Recommended tools for this procedure are a 10mm wrench, a 10mm nutdriver,
a 1/4 inch nutdriver, and a 3/16 inch nutdriver.
The replacement parts for this procedure are:
Receptacle, AC
Gasket, AC Receptacle
Seal, Housing, AC Receptacle
Bracket, AC Receptacle
Assembly, Cable, DataPort/Ground
Post, Equipotential
Shield, EMI
Nut, Hex, Metric, DIN 439
Screw, Lock, 4-40
CAUTION:
Use proper ESD grounding techniques when handling components.
Wear an antistatic wrist strap and use an ESD-protected workstation. Store the PWA
in an antistatic bag before placing it on any surface.
To replace the AC receptacle, refer to
Figure 7-11, AC Power Cord (220 VAC - Plum XLM)
and proceed as follows: