7.2
REPLACEMENT PROCEDURES
Technical Service Manual
7 - 29
430-00587-008 (Rev. 2/05)
3. Grasp the adhesive-backed insulator and remove it from the pole clamp backing
plate. Using a mild solvent, clean the pole clamp backing plate.
4. Using a 5/16 inch nutdriver, remove the two hex head screws securing the pole
clamp backing plate to the pole clamp extrusion. Remove the pole clamp backing
plate and pole clamp extrusion from the rear enclosure assembly.
5. Install the replacement pole clamp extrusion into the rear enclosure assembly.
6. Install the replacement pole clamp backing plate against the pole clamp extrusion.
Using a 5/16 inch nutdriver, install and tighten the two hex head screws.
7. Remove the adhesive backing from the replacement insulator. Completely cover the
pole clamp backing plate with the replacement insulator. Press firmly to adhere the
insulator to the backing plate.
CAUTION:
Make sure the insulator covers the entire backing plate. If the
backing plate is exposed, the power supply PWA may be damaged when power
is applied to the infusion system.
8. Install the power supply PWA in the exact reverse order of removal.
9. Join the front enclosure assembly and rear enclosure assembly in the exact reverse
order of separation.
Note:
(XLM with DataPort) When joining the front and rear enclosures, do not
pinch the DataPort/MCU cable between the two enclosures.
10. Connect the infusion system to AC (mains) power.
To verify successful pole clamp extrusion, backing plate, and insulator replacement,
perform the PVT in