SECTION 6 TROUBLESHOOTING
430-00587-008 (Rev. 2/05)
6 - 10
Plum XL Series
81-1
Power supply PWA
signals HKDC and
DHKDC do not match
Power supply PWA failure
Replace power supply
PWA
Failure in conditioning
circuit on MCU PWA
Replace MCU PWA
81-2
Power supply PWA
signal HKDC out of
tolerance
Power supply PWA failure
Replace power supply
PWA
Failure of conditioning
circuit on MCU PWA
Replace MCU PWA
81-3
Power supply PWA
signal VMOT out of
tolerance when AC
power is applied
Power supply PWA failure
Replace power supply
PWA
81-4
Motor voltage drops
too much when motor
is energized
Power supply PWA failure
Replace power supply
PWA
If condition recurs,
Motor drawing excessive
current
90-1
Calibration data in
EEPROM checksum
failure
EEPROM internal failure
EEPROM decode circuitry
failure
94-1
Control knob signal
seen when disabled
Control knob circuitry
failure
Replace display PWA
94-2
Illegal control knob
signal seen
Control knob circuitry
failure
Replace display PWA
94-4
Reed switch does not
match control knob signal
Control knob circuitry
failure
Replace display PWA
Reed switch failure
95-1
Primary valve pin
not moving
Pin detect circuitry failure
Replace mechanism
assembly
Valve pin not present
Valve pin not moving
95-2
Secondary valve pin
not moving
Pin detect circuitry failure
Replace mechanism
assembly
Valve pin not present
Valve pin not moving
96-1
UART test failure
during operation
UART failure
Replace MCU PWA
99-1
through
99-6
Failure of one
of the internal
software self-tests
MCU internal failure
Replace MCU PWA
Table 6-2.
Error Codes Requiring Technical Service
Error
Code
Malfunction
Description
Possible Cause
Corrective Action