6.3
TROUBLESHOOTING PROCEDURES
Technical Service Manual
6 - 15
430-00587-008 (Rev. 2/05)
Delivery accuracy
test
Set not properly primed
Reprime set
Damaged or faulty set
Prime using new set
Defective mechanism assembly
Replace mechanism assembly
Empty container/
air-in-line alarm test
Defective special cassette
Replace special cassette
Dirty bubble sensors
Clean bubble sensors
Defective bubble sensor PWA
Replace mechanism assembly
Defective sensor PWA
Replace mechanism assembly
Electrical safety test
Insufficient ground connection
Attach lead to T point screw
on rear of infusion system
Defective AC (mains) cordset
Replace AC (mains) cordset
Defective power supply PWA
Replace power supply PWA
Battery charger
current test
(optional)
Blown fuse
Replace fuse
Defective AC (mains) cordset
Replace AC (mains) cordset
Defective power supply PWA
Replace power supply PWA
DataPort connection/
ground continuity test
(optional)
Insufficient ground connection
from DataPort connector to
ground screw
Attach ground wire from DataPort
connector to ground screw
Defective DataPort/MCU cable
assembly
Replace DataPort/MCU cable
assembly
DataPort/MCU cable not properly
seated on MCU PWA
Reseat DataPort/MCU cable
assembly to MCU PWA
Table 6-6.
PVT Troubleshooting
Test Failures
Possible Causes
Corrective Actions