8.2
LIFECARE XL AND XLM
Technical Service Manual
8 - 3
430-00587-008 (Rev. 2/05)
Battery:
One 8 V, sealed, rechargeable battery, internal to the
infusion system. Accessible for ease of field replacement,
with color-coded leads and polarized connector
Battery Operating Time
(XL/XLM):
With a new, fully charged battery, the infusion system
provides approximately 1000 mL delivery volume or eight
hours of operation, whichever occurs first. The infusion
system displays a low battery alarm approximately 30
minutes prior to infusion system shutdown when pumping
at 125 mL/hr. If a low battery alarm occurs, immediately
connect the infusion system to mains power
Note:
Gradual degradation over extended periods of
use decreases the operational capacity of the battery.
Typical battery life is three years. A yearly check is
recommended to verify performance. When capacity
drops to an unacceptable level, replace the battery.
Battery replacement must be performed by qualified
technical personnel
Battery Operating Time
(XLM with DataPort):
With DataPort circuitry disabled, a fully charged new battery
provides eight hours of operation, or 1000 mL total volume
delivered, whichever occurs first. With DataPort enabled, a
fully charged new battery provides approximately five hours
of operation, or 625 mL total volume delivered, whichever
occurs first. The infusion system displays a low battery
alarm approximately 30 minutes prior to infusion system
shutdown when pumping at 125 mL/hr. If a low battery
alarm occurs, immediately connect the infusion system to
mains power
Note:
Do not connect DataPort when infusing
Note:
Gradual degradation over extended periods of
use decreases the operational capacity of the battery.
Typical battery life is three years. A yearly check is
recommended to verify performance. When capacity
drops to an unacceptable level, replace the battery.
Battery replacement must be performed by qualified
technical personnel
Battery Recharge:
The battery recharges when the infusion system is
connected to mains power. If the infusion system is
operating at 125 mL/hr, a full recharge takes approximately
16 hours. Recharge takes approximately eight hours if the
infusion system is turned to the OFF/CHARGE position
Battery Charge
Retention:
A fully charged battery retains at least 20 percent of its
capacity after six months of storage at below 35
°
C (95
°
F)
in an infusion system not connected to mains power
ENVIRONMENT
Operating Temperature:
0
°
to 40
°
C (50
°
to 104
°
F) 10 percent to 90 percent relative
humidity
Shipping and Storage:
-20
°
to 60
°
C (-4
°
to 140
°
F) 10 percent to 90 percent relative
humidity