5.2
PERFORMANCE VERIFICATION TEST
Technical Service Manual
5 - 7
430-00587-008 (Rev. 2/05)
CAUTION:
Do not place the infusion system in service if the self test fails.
To perform the self test, refer to
Figure 5-3, Plum XL LCD Test Screens
, then proceed as follows:
1. Connect the infusion system AC (mains) power cord to a grounded AC (mains) outlet
and confirm the AC (mains) power icon (next to the OFF/CHARGE setting)
illuminates.
2. Open the door assembly (cassette door) by lifting up on the cassette door handle.
3. Hold a primed cassette by its handle and insert the cassette into the cassette door
guides. Do not force the cassette into position.
4. Close the cassette door handle to lock the cassette in place.
5. Turn the control knob to SET RATE to initiate the self test.
6. Verify the following screens display: the LCD test screen; four backward Cs
(approximately two seconds); set rate screen.
Note:
If the LCD test screen does not match
contact Hospira.
Note:
If an alarm condition occurs during the self test, turn the control knob
to OFF/CHARGE and repeat Step 5 and Step 6. If the alarm condition recurs,
the message and take corrective action
(see Section 6, Troubleshooting)
. Repeat
the self test. If the alarm condition recurs, remove the infusion system from
service and
7. Disconnect the infusion system from AC (mains) power and confirm BATTERY
displays on the LCD screen.
8. Turn the control knob to OFF/CHARGE and remove the administration set.
9. To allow the battery to charge fully, connect the infusion system to AC (mains) power
for a minimum of eight hours with the control knob in the OFF/CHARGE position.
Confirm the AC (Mains) power icon illuminates.
Figure 5-3.
Plum XL LCD Test Screens
KVO 2 1 mL=mL mL/H,mL?
04K01024
ENGLISH LANGUAGE
2 1 mL/H
ML
AIR IN LINE
EMPTY
LOW BATTERY
KVO
VTBI COMPLETE
DOOR/CASSETTE
TURN TO RUN CHECK SETTINGS OCCLUSION
SECONDARY PRIMARY BACKPRIMING
SET
RATE
VTBI
PIGGYBACK ML/HR
TOTAL VOLUME
DELIVERED
OK
ICON BASED