SECTION 7 REPLACEABLE PARTS AND REPAIRS
430-00587-008 (Rev. 2/05)
7 - 20
Plum XL Series
9. Connect the AC (mains) power cord wire to J17
10. Connect the AC (mains) power cord wire to J16
11. Connect the ground wire with the grounding lug to the ground screw. Using a 1/4
inch nutdriver, replace and tighten the two hex head nuts to the ground screw.
12. Reassemble the device in the exact reverse order of disassembly.
Note:
(XLM with DataPort) When joining the front and rear enclosures, do not
pinch the DataPort/MCU cable between the two enclosures.
13. Connect the infusion system to AC (mains) power.
To verify successful AC (mains) power cord, strain relief bushing, and Velcro strap
replacement, perform the PVT in
Figure 7-8.
AC Power Cord (115 VAC - Plum XL)
REAR ENCLOSURE
AC POWER
CORD
EMI SHIELD
POWER
SUPPLY PWA
04K01044
STRAIN RELIEF
BUSHING
6-32 KEP NUT (3)
FLAT WASHER (3)
GROUND WIRE (2)
6-32 x 3/4 HEX
HEAD SCREW