6.2
ALARM MESSAGES AND ERROR CODES
Technical Service Manual
6 - 5
430-00587-008 (Rev. 2/05)
18-2
Display blank
Infusion system
shutdown one
minute after
discharged
battery alarm
Fully
discharged
battery
Connect to AC
power, turn
control knob to
OFF/CHARGE,
or replace
the battery
19-1
DOOR
DOOR/CASSETTE
(XL only
or
Door open
Cassette
door open
Turn control
knob to
OFF/CHARGE,
or close
cassette door
Cassette not
seated properly
Reseat cassette
Table 6-1.
Operational Alarm Messages and Corrective Actions
Error
Code
Alarm Message
Description
Possible Cause
Corrective
Action