7.2
REPLACEMENT PROCEDURES
Technical Service Manual
7 - 21
430-00587-008 (Rev. 2/05)
Figure 7-9.
AC Power Cord (115 VAC - Plum XLM)
AC (MAINS) POWER CORD REPLACEMENT (INTERNATIONAL)
No tools are required for this procedure.
The replacement part for this procedure is:
Cordset, AC Power, Hospital, International
To replace the AC (mains) power cord, refer to
Figure 7-10, AC Power Cord (220 VAC - Plum
Figure 7-11, AC Power Cord (220 VAC - Plum XLM)
, then disconnect the power cord
from the rear of the infusion system and connect the replacement power cord.
Replacement of the AC (mains) power cord is a routine maintenance procedure and no
verification procedure is normally required. However, if the infusion system may have been
damaged during this procedure, perform the PVT in
EMI
SHIELD
REAR ENCLOSURE
AC POWER
CORD
DATAPORT/GROUND
CABLE ASSEMBLY
04K01045
STRAIN
RELIEF
BUSHING
POWER
SUPPLY
PWA
FLAT WASHER (2)
6-32 KEP NUT (4)
GROUND WIRE (2)
6-32 x 1 HEX
HEAD SCREW