SECTION 7 REPLACEABLE PARTS AND REPAIRS
430-00587-008 (Rev. 2/05)
7 - 34
Plum XL Series
10. Insert the pointed end of the cable tie through the corner locating hole of the MCU
PWA and loop back around the edge of the PWA.
11. Insert the pointed end of the cable tie into the locking head.
12. Using needle nose pliers, carefully tighten the cable tie until tight.
13. Verify the cable tie is tight and cut away any excess using the diagonal cutters.
14. Reassemble the device in the exact reverse order of disassembly.
Note:
(XLM with DataPort) When joining the front and rear enclosures, do not
pinch the DataPort/MCU cable between the two enclosures.
15. Connect the infusion system to AC (mains) power.
To verify successful MCU PWA replacement, perform the PVT in
Figure 7-16.
Exploded View of the Main Chassis Assembly
04K01011
OPENER HANDLE
ASSEMBLY
MECHANISM
ASSEMBLY
MCU PWA
6-32 x 1/2
HEX HEAD
SCREW
MAIN CHASSIS
ASSEMBLY
SENSOR/MCU
CABLE ASSEMBLY