SECTION 5 MAINTENANCE AND SERVICE TESTS
430-00587-008 (Rev. 2/05)
5 - 6
Plum XL Series
Inspect the infusion system periodically for signs of defects such as worn accessories or
damaged cables. Also, inspect the infusion system after repair or during cleaning. Replace
any damaged or defective external parts.
Inspect the following areas for missing and damaged parts:
Note:
Do not connect DataPort when infusing.
To set up the infusion system for the PVT, proceed as follows:
1. Confirm the infusion system and appropriate accessories are fully assembled.
2. Hang two sterile water containers at a height of 18 to 24 inches (46 to 60 cm) above
the pumping chamber of the infusion system.
3. Connect the infusion system to AC (mains) power. Conduct all tests with the
infusion system connected to AC (mains) power unless otherwise specified.
4. Verify the lockout switch is in the UNLOCKED (down) position (XLM only).
-
Labels
-
Front panel label
-
AC power cord
-
Control knob and all external screws
-
Velcro
®
retainer strap
-
Pole clamp assembly
-
Rubber foot pads
-
Front and rear enclosures
-
Door assembly, shield, and handle
-
Battery access cover
-
Cassette guide spring and roller
-
LCD screen
-
Valve pins, plunger, bubble
detectors, and locator pin
-
DataPort connector (XLM with DataPort)
WARNING:
A PATIENT SHOULD NEVER BE CONNECTED TO THE
INFUSION SYSTEM DURING SYSTEM TESTING.