APPENDIX
430-00587-008 (Rev. 1/05)
A - 2
Plum XL Series
Table A-2, Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
details
guidance for the electromagnetic environment for XL series infusion pumps
Table A-2.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6kV Contact
±8kV Air
±8kV Contact
±15kV Air
(See Note 2)
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2kV for power supply
lines
±1kV for input/output
lines
±2kV for power
supply lines
±1kV for input/output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment
Surge
IEC 61000-4-5
±1kV differential mode
±2kV common mode
±1kV differential
mode
±2kV common mode
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines IEC
61000-4-11
<5%U
r
(>95% dip in U
r
)
for 0.5 cycle
40%U
r
(60% dip in U
r
)
for 5 cycles
70%U
r
(30% dip in U
r
)
for 25 cycles
5%U
r
(>95% dip in U
r
)
for 5 seconds
<5%U
r
(>95% dip in
U
r
) for 0.5 cycle
40%U
r
(60% dip in U
r
)
for 5 cycles
70%U
r
(30% dip in U
r
)
for 25 cycles
5%U
r
(>95% dip in U
r
)
for 5 seconds
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the LifeCare
XLM with DataPort
requires continued
operation during power
mains interruptions, it
is recommended that
the LifeCare XLM with
DataPort be powered
from an uninterruptible
AC Mains power
supply or the battery.
Power
Frequency (50/
60 Hz) magnetic
field IEC
61000-4-8
3A/m
400A/m
(See Note 3)
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Note 1:
U
r
is the AC Mains voltage prior to application of the test level.
Note 2:
Compliance levels tested to IEC 60601-2-24 requirements, which are more stringent
than IEC 61000-4-2.
Note 3:
Compliance levels tested to IEC 60601-2-24 requirements, which are more stringent
than IEC 61000-4-8.