Cenova Image Analytics Server 3.0 Installation & Service Manual
Chapter 3: Installing the System
MAN-05205 Revision 003
Page 43
•
SC – TOMO.
This Secondary Capture (SC) image format provides proprietary
ImageChecker 3D Calc CAD results for display on Hologic SecurView workstations
only.
•
Printer – CAD.
This format provides a printout of the four case-processed views
with ImageChecker CAD (2D) results similar to the SC - CAD layout.
Note
The Cenova server does not support DICOM printing.
•
MG – Enhanced.
This DICOM Digital Mammography X-Ray (MG) Image - For
Presentation format provides DigitalNow HD images. The DigitalNow HD results
are updated copies of the DigitalNow images that are received as input, processed
according to the HD configuration settings (e.g., sizing, LUTs, W/L values).
Note
DigitalNow HD can produce output images at any resolution from 50 to 200
microns. Be sure to review the DigitalNow HD
on page 64 guidelines
to understand output image resolution.
To determine the appropriate Edit DICOM Mappings format, consult the documentation
provided by the workstation vendor. For a discussion of the licensable output features,
see
on page 49.