Cenova Image Analytics Server 3.0 Installation & Service Manual
Chapter 5: Service Procedures
MAN-05205 Revision 003
Page 97
To install the updates:
1.
Log into Windows as
Administrator
.
2.
Load the Cenova System Updates removable media.
3.
Navigate to the drive and double-click
start.cmd
. The Cenova System Updates
window appears.
4.
Click
Start
. The Cenova System Updates installer copies files to disk and configures
application settings.
Note
If during intermediate reboots the Administrator password is rejected at login,
try again and leave the password blank.
5.
When prompted to restart Windows, click
Yes
.
6.
Remove the Cenova System Updates removable media and store in a safe location.
7.
Log into Windows as
Administrator
.
8.
Load the Hologic OS Updates removable media (x64 for Windows 7).
9.
Navigate to the drive and double-click
SystemUpdater.exe
. The Hologic OS Updates
window appears.
10.
Click
Start
. The Hologic OS Updates installer copies files to disk and updates system
files. During this process, which can take up to 20-30 minutes, the installer
automatically reboots the system
once
.
11.
When prompted for the Administrator password, enter the password and click
Accept
.
12.
When prompted to restart Windows, click
Yes
.
13.
Remove the Hologic OS Updates removable media and store in a safe location.
14.
Log into Windows as
Administrator
.
15.
Verify the video drivers are correctly installed (ImageChecker 3D Calc CAD
application servers only).
a.
Select
Start > Control Panel
.
b.
Select
System
.
c.
Select
Device Manager
from the left menu. The Device Manager window opens.
d.
Double-click
Display Adapters
.
e.
Confirm that the device drivers are listed.
Note
If the video drivers were not installed, rerun the system updates.
To install the server software: