Cenova Image Analytics Server 3.0 Installation & Service Manual
Chapter 5: Service Procedures
MAN-05205 Revision 003
Page 95
To install the server software:
1.
Install the Cenova server software by running the installer executable file (see
Installing the Server Software
on page 28 for instructions).
To restore the configuration settings:
1.
Restore the Cenova Control Panel settings (see
Backing Up and Restoring the Server
on page 77), then step through all configuration screens to verify the settings
are correct.
2.
Restart the Hologic Connect service (see
Configuring the Hologic Connect Service
page 65).
3.
Run the Self Test (see
on page 81).
4.
Verify system operation.
Note
If something goes wrong during the Cenova application reinstallation process, you
need to reimage the system. Follow instructions in
on page 96.)
Note
If DICOM 6000 Interface is installed on the Cenova server, refer to the DICOM 6000
Interface Installation & Service Manual to reinstall and configure DICOM 6000 Interface
if necessary.