87
________________________________________________________________________________________________________________
6.15 Commands Specific to the 3511-50
________________________________________________________________________________________________________________
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
Unused Unused Unused Unused Unused Overrun
error
Framing
error
Parity
error
RS-232C communication condition errors register
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
CEM
SOF
SUF
MOF
MUF
IDX
EOM
Unused
Event status register 0 (ESR0)
:ERRor?
Queries RS-232C communication condition errors.
:ESR0?
Queries event status register 0.
Syntax
:ERRor?
Function
Returns the value of RS-232C communication condition errors as a numerical value in
NR1 format from 0 to 7, and then clears RS-232C communication condition errors.
No header is prefixed to the response message.
<data>
Numerical data in NR1 format between 0 and 7
Example
Response
4
An overrun error has occurred.
Error
If the response message is longer than 300 bytes, a query error is generated.
Syntax
ESR0?
Function
Returns the value of event status register 0 (ESR0) as a numerical value in NR1 format
between 0 and 255, and then clears event status register 0.
No header is prefixed to the response message.
Example
Response
4
Bit 2 of ESR0 has been set to 1.
Error
If the response message is longer than 300 bytes, a query error is generated.