25
________________________________________________________________________________________________________________
3.8 Short Circuit Compensation
________________________________________________________________________________________________________________
3.8.2 Short Circuit Compensation Error
3.8.3 Canceling Short Circuit Compensation
The following are possible causes:
1. The circuit across HIGH and LOW terminals is not short.
2. The test cable is not connected correctly.
3. The impedance across the terminals exceeds 1 k
Ω
.
Procedures
(1) Press
with short circuit compensation ON. Short circuit
compensation is canceled. The value compensated for is not saved.
(2) The short circuit compensation will remain invalid until the
compensation data is deleted and another short circuit compensation is
performed.