3.7 Open Circuit Compensation
________________________________________________________________________________________________________________
NOTE
3.7.2 Open Circuit Compensation Error
3.7.3 Canceling Open Circuit Compensation
If an error occurs during open circuit compensation, the unit beeps to notify
you of the error. An error message appears, as shown, and compensation is
aborted.
Press
to return the unit to normal test mode.
If an error occurs, when you return to normal test mode after aborting open
circuit compensation, the measurement value is not compensated for.
The following are possible causes:
1. The circuit across HIGH and LOW terminals is not open.
2. The test cable is not connected correctly.
3. The impedance across the terminals does not reach 1 k
Ω
.
Procedures
(1) Press
with open circuit compensation ON. Open circuit
compensation is canceled. The value compensated for is not saved.
(2) The open circuit compensation will remain invalid until the
compensation data is deleted and another open circuit compensation is
performed.