4.2 Panel Save Function
The current test conditions may be saved to internal memory, up to a
maximum of 99 panels (or combinations) of test conditions.
All the set conditions are saved when panel save is performed, including
comparator upper and lower limits and OPEN/SHORT compensation values.
You can read saved measurement conditions later with the panel load
function.
(1) Press
repeatedly to cycle through available settings: panel load, panel
save and normal test (or comparator test) modes. For more information on
panel load mode, see Section 4.3, "Panel Load Function."
(2) Enter panel save mode and specify the panel number to save. In panel save
mode, only unsaved panel numbers are displayed. (The panel number at
shipment is "01.")
If all panels are used for the saving, when the panel mode is entered, "01"
appears.
To overwrite a panel already in use, choose the panel number and correct.
Use the COUNT setting key to specify a panel number.
keys increase/decrease the value.
move the cursor to a digit to be set (up to two digits).
The LED for the digit under setting flashes.
The set values must fall between 01 and 99.
(3) After specifying a panel number, press
. The 3511-50 saves the test
conditions and reverts to normal or comparator test mode.
After specifying a panel number in panel save mode, press
instead of
. The 3511-50 reverts to normal or comparator test mode without
performing the panel save.
Under normal conditions of use, the average life of the backup battery for
the internal memory is about 6 years.
If the internal battery becomes exhausted, it is no longer possible to save
the test conditions. You should have the battery changed by an approved
HIOKI
service facility (which is chargeable).