18
________________________________________________________________________________________________________________
3.6 Setting the Ranging
________________________________________________________________________________________________________________
Test range
Z
R
Frequency
Common to
120 Hz and 1 kHz
Common to
120 Hz and 1 kHz
Range number
Range
Display range
Range
Display range
10
100 M
Ω
200.00 M
Ω
100 M
Ω
200.00 M
Ω
009.00 M
Ω
009.10 M
Ω
9
10 M
Ω
9.9999 M
Ω
10 M
Ω
9.9000 M
Ω
0.9000 M
Ω
0.9100 M
Ω
8
1 M
Ω
999.99 k
Ω
1 M
Ω
990.00 k
Ω
090.00 k
Ω
091.00 k
Ω
7
100 k
Ω
99.999 k
Ω
100 k
Ω
99.000 k
Ω
09.000 k
Ω
09.100 k
Ω
6
10 k
Ω
9.9999 k
Ω
10 k
Ω
9.9000 k
Ω
0.9000 k
Ω
0.9100 k
Ω
5
1 k
Ω
999.99
Ω
1 k
Ω
990.00
Ω
090.00
Ω
091.00
Ω
4
100
Ω
99.999
Ω
100
Ω
99.000
Ω
09.000
Ω
09.100
Ω
3
10
Ω
9.9999
Ω
10
Ω
9.9000
Ω
0.9000
Ω
0.9100
Ω
2
1
Ω
0.9999
Ω
1
Ω
0.9900
Ω
0.0900
Ω
0.0900
Ω
1
100 m
Ω
0.0999
Ω
100 m
Ω
0.0990
Ω
0.0100
Ω
0.0110
Ω
To measure a parameter other than Z - i.e., C, L, or R - determine an
appropriate range based on impedance. The following table presents
appropriate ranges for C and L when D is equal to or smaller than 0.1 and