4.1 Comparator Function
________________________________________________________________________________________________________________
4.1.1 Operation Sequence
Normal test mode
Second parameter lower
limit setting mode
Second parameter
upper limit setting mode
First parameter lower
limit setting mode
First parameter upper
limit setting mode
Comparator test mode
4.1.2 Setting the Upper and Lower Limits
Conditions being
set
Status display
Comparator judgment
display
First parameter
upper limit
"M-HSET" lights.
"HI" lights to indicate the judgment
result for the first parameter.
First parameter
lower limit
"M-LSET" lights.
"LO" lights to indicate the judgment
result for the first parameter.
Second parameter
upper limit
"S-HSET" lights.
"HI" lights to indicate the judgment
result for the second parameter.
Second parameter
lower limit
"S-LSET" lights.
"LO" lights to indicate the judgment
result for the second parameter.
Set this among test conditions for executing a comparator.
Press
.
Automatically switches the range setting to HOLD.
Use the COUNT setting key to set the upper limit.
Press
to acknowledge and save the set value; then set the
lower limit.
Use the COUNT setting key to set the lower limit.
Press
to acknowledge and save the set value; then set the
second parameter upper limit.
Use the COUNT setting key to set the upper limit.
Press
to acknowledge and save the set value; then set the
lower limit.
Use the COUNT setting key to set the upper limit.
Press
to acknowledge and save the set value.
If all upper and lower limits are OFF at this point, the system
switches to normal test mode. If setting upper and lower limits is
valid, the system switches to comparator test mode.
Press
, and the system switches to normal test mode.
(1) Press
to enter a desired setting mode, for either the upper or lower
limit.
The current upper and lower limit settings are displayed in the status display
next to
, and in the comparator judgment display.
Set values are displayed at the following locations.
First parameter upper and lower limits : First parameter measurement
display
Second parameter upper and lower limits : Second parameter measurement
display
If the upper or lower limit value is set to OFF, the displayed setting will be
"-----." Upper or lower limit values set to OFF are not used for comparison
with a measurement.