6.15 Commands Specific to the 3511-50
________________________________________________________________________________________________________________
:CIRCuit?
Queries the equivalent circuit mode.
:CIRCuit:Auto
Sets the automatic equivalent circuit mode.
:CIRCuit:Auto?
Queries the automatic equivalent circuit mode.
Syntax
:BEEPer:KEY?
<data>
SER/PAR (character data)
Function
Returns the current equivalent circuit mode setting as character data.
SER: Series equivalent circuit mode
PAR: Parallel equivalent circuit mode
Example
Response
If headers are on
:CIRCUIT SER
If headers are off
SER
Error
If the response message is longer than 300 bytes, a query error is generated.
Syntax
:CIRCuit:AUTO
<data>
<data>
ON/OFF (character data)
Function
Switches between automatic and manual setting of equivalent circuit mode.
ON: Switches the automatic setting.
OFF: Switches the manual setting.
Example
Transmission
":CIRCuit:AUTO ON"
The equivalent circuit mode is switched to automatic selection (auto-ranging).
Error
If <data> is other than character data described above, an execution error occurs.
Executing this command while the open or short circuit compensation is performed
generates an execution error.
Syntax
:CIRCuit:AUTO?
<data>
ON/OFF
Function
Returns whether the equivalent circuit mode is automatically set as character data.
Example
Response
If headers are on
":CIRCUIT:AUTO ON"
If headers are off
"ON"
Error
If the response message is longer than 300 bytes, a query error is generated.