13
________________________________________________________________________________________________________________
3.4 Setting the Testing Speed
________________________________________________________________________________________________________________
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
Test frequency
FAST
NORM
SLOW
120 Hz
13 ms
90 ms
400 ms
1 kHz
5 ms
60 ms
300 ms
NOTE
3.4 Setting the Testing Speed
For testing speed, select from one of the following three levels, depending on
the specific purpose. The slower the testing speed is, the more accurate are
the results.
FAST
: Low accuracy testing is performed at high speed.
NORM :
The speed used for normal testing
SLOW :
High accuracy testing is performed slowly.
Procedure
Press
repeatedly to cycle through available settings: FAST, NORM, and
SLOW. Use the LED lamp to the left of the key to check on a set testing
speed.
Testing speed
The testing time values above are for reference only, valid when Z-
θ
is
displayed and varying according to testing parameter setting conditions,
range mode, OPEN/SHORT compensation, and other conditions.
Summary of Contents for 3511-50
Page 1: ...INSTRUCTION MANUAL For 3511 50 LCR HiTESTER...
Page 2: ......
Page 137: ......