3.2 Setting the Test Frequency
________________________________________________________________________________________________________________
3.2 Setting the Test Frequency
3.3 Setting the Test Signal Level
For the 3511-50's test frequency, you may specify either 120 Hz or 1 kHz.
Procedure
Press
to toggle between 120 Hz and 1 kHz. Use the LED lamp to the
left of the key to check on a set test frequency.
For the 3511-50's test signal level, you may specify 1 V, 500 mV or 50 mV.
Procedure
Press
repeatedly to cycle through available settings: 1 V, 500 mV, and
50 mV. Use the LED lamp to the left of the key to check on a set test
signal level.
Precautions for changing test signal levels
The value of the test signal level may change according to the sample which
is being tested. During measurement, keep in mind that inductance is often
current-dependent.