4.5 Various Settings Made After Switching on Power
________________________________________________________________________________________________________________
4.5.4 Executing System Reboot
Display in the first parameter
measurement display
Setting
Display in the second
parameter measurement
display
No system reboot
System reboot
The conditions in force at shipment are as follows:
Test mode
Normal test mode
Test parameters
|Z|-
θ
Test frequency
1 kHz
Test signal level
1 V
Test speed
NORM
Equivalent circuit mode
AUTO
Test range
AUTO
Open circuit
compensation
OFF
Short circuit
compensation
OFF
Trigger mode
INT (Internal trigger)
Key lock function
OFF
Beep sound setting
OFF (when a comparator judgment is
made)
Panel save
All contents clear
Comparator
Upper and lower limits for the first and
second parameters both OFF
All test conditions are initialized to the conditions in force at shipment.
All measurement conditions include panel-saved conditions.
Procedures
Restart the 3511-50 while pressing
. All displays will light, and in
approximately 1.5 s, the Section switches to the version information screen.
You can release the
key when this occurs.
The 3511-50 displays version information, "rESEt" in the first parameter
measurement display and "no" in the second parameter measurement display.
Press the COUNT setting key to switch the display in the second parameter
measurement display between "no" and "YES." Press
in either state.
Pressing
in the "YES" state initializes all test conditions to the
conditions in force at shipment and switches the screen to "Interface setting."